Research of Reference Range of 6 Reticulocyte Parameters

ZHANG Qing-fang,AI Dong-qin,CHEN Ya-bao(Taizhou People’s Hospital,Taizhou Jiangsu 225300,China)  
Objective To establish the reference range of 6 reticulocyte parameters with Sysmex XT-2000i henmatology analyzer.Methods Detect reticulocyte ratic(RET%),reticulocyte count(RET#),low fluorescence ratio(LRF),middle fluorescence ratio(MRF),high fluorescence ratio(HRF),immature retics fraction(IRF),using Sysmex XT2000i henmatology analyzer.Results RET%:male was 0.721±0.202,female was 0.714±0.232,there was no significant difference between them(P0.05);unite RET% 0.718±0.217.RET#:male was 35.51±10.34,female was 31.42±10.53,there was significant difference(P 0.05).IRF: male was4.21±2.3,femalewas4.61±2.26,unite IRF4.41±2.28.LFR%:male was95.79 ±2.3,female was95.39 ±2.26,unite LFR%95.59 ±2.28.Conclusion The results between the research of reference range of 6 reticulocyte parameters and the records of the document have some differences.Different labs should establish their own reference range according to their own instruments.