Quality Control Inspection of Radiotherapy Equipment Based on Image Guided Radiation Therapy System

CHEN Li-jun;GU Xiao-hua;YANG Liu-qin;Department of Radiotherapy,Xinxiang Central Hospital;  
Objective To discuss the possibility of periodical inspection for partial quality control of linear accelerator by combining homemade verification tool and the image guided radiation therapy system. Methods Putting verification plate according to different purposes under conditions that beam central axis was perpendicular to verification plate. Different exposure technique and image acquisition mode were used to capture images of the quality control verification of accelerator. Results The difference between the cross image center and beam axis or reference image center is no more than 2 mm. Conclusion The method is accurate, economical and practical, and also easy to operate in the application of quality control project inspection of the linear accelerator.
【CateGory Index】: R197.39
Download(CAJ format) Download(PDF format)
CAJViewer7.0 supports all the CNKI file formats; AdobeReader only supports the PDF format.
【Citations】
Chinese Journal Full-text Database 5 Hits
1 Gu Xiaohua;Yang Liuqin;Department of Radiotherapy,Xinxiang Central Hospital;;Thrice exposure technology applied for electron beam boost at the nape of the neck in radiotherapy[J];现代肿瘤医学;2015-24
2 GU Xiao-hua;YANG Liu-qin;Department of Radiation Oncology, Xinxiang Central Hospital;;Application of IP plate in the quality assurance inspection of radiotherapy[J];中国医学装备;2015-02
3 Yu De-quan,Qu Huan-min,Shao Qiu-ju,Liang Jun,Gao Hong,(Department of Radiotherapy,Tang Du Hospital,The Fourth Military Medical University,Xi'an Shanxi 710038,China);Electron Portal Image Device in Accelerator Radiation Field Application in The Measurement of The Discussed[J];中国医学物理学杂志;2013-02
4 CHEN Jian-gong (Medical Room,Fujian Institute of Measurement Science and Technology,Fuzhou Fujian 350003,China);Detection,Ajustment and Improvement Study of Coincidence Degree between Radiation Field and Optical Field[J];中国医疗设备;2009-10
5 DENG Xiao-wu;Quality Control and Quality Assurance for Radiotherapy[J];中国肿瘤;2008-08
【Co-citations】
Chinese Journal Full-text Database 10 Hits
1 GU Xiao-hua;YANG Liu-qin;Department of Radiation Oncology, Xinxiang Central Hospital;;Application of IP plate in the quality assurance inspection of radiotherapy[J];中国医学装备;2015-02
2 FU Hong-jie;Department of Radiotherapy 1st Hospital of Xinxiang;;Measurement analysis of the dose output stability of NMSR600 medical linear accelerator[J];中国医学装备;2014-12
3 LI Zhi-cong;LI Lu-jun;YU Hai-kun;MA Zi-ming;ZHOU Qing-hua;Department of Radiotherapy, Wuzhou Red Cross Hospital;;Accuracy detection of multileaf collimator movement by dose verification[J];医疗卫生装备;2014-09
4 ZHANG Hai-cheng;Department of Medical Engineering, People' Hospital, DaXing District;;Research on the principle and application of high voltage of the Elekta accelerator[J];中国医学装备;2014-09
5 Hu Rui;Wang Shi;Wu Jinchang;Shen Danqing;Wu Zhaoxia;Department of Radiation Oncology,Affiliated Suzhou Hospital of Nanjing Medical University;Key Laboratory of Particle& Radiation Imaging(Tsinghua University),Ministry of Education;;Probabilistic safety assessment method in the application of external beam radiotherapy process control management study[J];中华放射肿瘤学杂志;2014-05
6 YANG Weiqiang YANG Bo KONG Xudong ZHANG Jian Department of Tumor Radiotherapy,the Fourth Affiliated Hospital of Suzhou University,Jiangsu Province,Wuxi 214062,China;Quality control of intensity modulated radiotherapy for tumor[J];中国当代医药;2012-29
7 LIU Wen,MEI Shi-bing,XU Tao Medical Engineering Department,The Military General Hospital of Beijing PLA,Beijng 100700,China;The Design and Application of the Coherence Metrical Ruler of X-Ray Confine[J];中国医疗设备;2011-10
8 Zhang Guofeng,Tang Yan,Sa Ying,Ma Aifang Department of Hematology and Oncology,Karamay Centre Hospital,Karamay Xinjiang 834000,China;Daily Quality Assurance of Linac Radiation Field[J];中国医疗器械杂志;2011-05
9 MO Li, LIAO Fu-xi, XU Zhi-hai (Radiotherapy Center, the 303 rd Hospital of PLA, Nanning 530021, China);Development and Experiment of Manual Calibration for Siemens Multileaf Collimator[J];医疗卫生装备;2011-08
10 XU Feng,CHEN Wei-wei Medical Engineering Department,Peking University Third Hospital,Beijing 100191,China;The Brief Description of Medical Linear Accelerator Technology[J];中国医疗设备;2011-04
【Secondary Citations】
Chinese Journal Full-text Database 10 Hits
1 KE Qi(Department of Radiology,Shantou Central Hospital,Shantou 515011,China);Shooting and Assessing the Portfilm in Radiation Therapy[J];中国医药指南;2013-07
2 DONG Hai-bin Department of Medical Imaging, Jinan Military General Hospital, Jinan 250031, China.;Quality assurance and quality control of X-ray digital radiography[J];中国医学装备;2012-06
3 ;EPID在盆腔肿瘤放疗中的摆位误差分析[J];当代医学;2012-02
4 Zhang Guofeng,Tang Yan,Sa Ying,Ma Aifang Department of Hematology and Oncology,Karamay Centre Hospital,Karamay Xinjiang 834000,China;Daily Quality Assurance of Linac Radiation Field[J];中国医疗器械杂志;2011-05
5 LIU Peng,YU Wei-hai, XIAO Kai-min Equipment Department, Qingdao Haici Medical Group,Qingdao Shandong 266000,China;Quality Assurance and Quality Control of Medical Linear Accelerator[J];中国医疗设备;2010-09
6 ZENG Zi-li (Department of Tumor, Liutie Central Hospital, Liuzhou 545007, Guangxi Province, China);Quality Assurance and Quality Control of Medical Electron Linear Accelerator[J];医疗卫生装备;2010-09
7 YU De-quan,GAO Hong,SHAO Qiu-jiu,LIANG Yun,ZHANG Xin(Department of Radiotherapy,Tang Du Hospital,The Fourth Military Medical University,Xi'an Shannxi 710038,China);Application and Research of Two-dimensional Ion Chamber Array in Daily Quality Assurance of Accelerator[J];中国医学物理学杂志;2010-04
8 YU Da-hai,LIU Ling,REN Hong-rong,GUAN LuanDepartment of Radiation Oncology of Xuzhou Central Hospital,Xuzhou 221009,P.R.China;Clinical research of setup optimization in radiotherapy by EPID[J];中华肿瘤防治杂志;2010-06
9 CHEN Jian-gong (Medical Room,Fujian Institute of Measurement Science and Technology,Fuzhou Fujian 350003,China);Detection,Ajustment and Improvement Study of Coincidence Degree between Radiation Field and Optical Field[J];中国医疗设备;2009-10
10 LI Zhao-bin,LU Yao-hong,SUN Yi,HUANG Guo-feng,ZHANG Xiu-long,FU Shen(Department of Radiation Oncology,the 6th hospital of Shanghai Jiaotong University,Shanghai 200233,China);Application of Megavolt Computed Radiography for Portal Field Verification[J];中国医学物理学杂志;2009-05
©2006 Tsinghua Tongfang Knowledge Network Technology Co., Ltd.(Beijing)(TTKN) All rights reserved