Exploration of exposure dose optimization for pelvic examination in direct digital radiography with a phantom

WANG Peng-cheng *,ZHANG Fu-li, LI Shi-jun. *Department of Radiology, Taishan Medical College, Shandong 271000, China  
Objective To investigate the relationship between exposure dose and image quality for a direct digital radiography system and to determine the optimum exposure parameters for pelvic examination. Methods A contrast-detail phantom CDRAD2.0 was exposed to different doses and the IQF of phantom images were calculated. The optimum exposure parameters were determined by ANOVA and SNK analysis. The image of an anthropomorphic phantom taken with optimized exposure parameters was verified using CEC image criteria. Results The IQF of the images of CDRAD2.0 phantom was significantly different for different doses, as the exposure dose was greater than 0.61 mGy. The IQFs have no difference for different dose groups. The image quality between this optimized exposure dose and conventional exposure dose not was significantly different. Conclusion The image quality of direct digital radiography system is improved as the exposure dose increases, and as the exposure dose exceeds a certain threshold, the image quality will not change apparently. There is an optimum exposure dose and the optimized exposure dose for the researched phantom is 0.61 mGy.
【CateGory Index】: R144;
Download(CAJ format) Download(PDF format)
CAJViewer7.0 supports all the CNKI file formats; AdobeReader only supports the PDF format.
【References】
Chinese Journal Full-text Database 3 Hits
1 TIAN Lan, GONG Jianping, JIANG Zhen,et al.Department of Radiology,The Second Affiliated Hospital of Soochow University,Suzhou,Jiangsu Province 215004,P.R.China;Application of Low Dose Photography in Pelvic Digital Radiography[J];Journal of Clinical Radiology;2011-02
2 LIU Rui-hong1,QI Wei-wei2,FU hai-hong3,LU Yong-qin2(1.Radiology Deparment of Sino-Japan Friendship Hospital,Beijing 100029,China;2.Radiology Deparment of People's Hospital,Beijing Univeristy,Beijing 100044,China;3.Radiology Deparment of PUMC Hospital,Beijing 100730,China);Effect on DR Pelvimetry with Automatic Exposure Control by Different kVp[J];China Medical Equipment;2008-04
3 LI Juan,HU Zhi-hai,HE Wen(Department of Radiology,Beijing Friendship Hospital,Capital Medical University,Beijing 100050,China);Quantitative evaluation of density resolution of DR images with CDRAD 2.0 phantom[J];Chinese Journal of Medical Imaging Technology;2011-10
【Co-references】
Chinese Journal Full-text Database 10 Hits
1 CAO Guo-quan,ZHOU Xiao-jun,SUN Hou-zhang,et al.Department of Radiology,the First Affiliated Hospital of Wenzhou Medical College,Wenzhou 325000,P.R.China;A comparative experimental study of CR vs DR in chest radiography parameter[J];Radiologic Practice;2010-06
2 WANG Wu,CAO Guo-quan,ZHU Wei The First Affiliated Hospital of Wenzhou Medical College (Wenzhou,325000);Comparison between Two Digital Imaging Detector System:Image Qulity and Radiation Dose[J];Progress in Biomedical Engineering;2009-02
3 YANG Dong 1,LIU Miao Fang 2,PAN Di 1(1.First Hospital Affiliated to Guangzhou Medical College,Guangzhou Guangdong 510120,China;2.Guangzhou Institute of Respiratory Disease,Guangzhou Guangdong 510120,China);The principle and maintenance of DR imaging system based on flat panel technique[J];Information of Medical Equipment;2006-12
4 LIU Rui-hong1,QI Wei-wei2,FU hai-hong3,LU Yong-qin2(1.Radiology Deparment of Sino-Japan Friendship Hospital,Beijing 100029,China;2.Radiology Deparment of People's Hospital,Beijing Univeristy,Beijing 100044,China;3.Radiology Deparment of PUMC Hospital,Beijing 100730,China);Effect on DR Pelvimetry with Automatic Exposure Control by Different kVp[J];China Medical Equipment;2008-04
5 ZHANG Xi-yong(The First Affiliated Hospital to Nanjing Medical University, People's Hospital of Jiangsu Province,Nanjing 210029,China);Analysis of Exposure Dose Between CR System and Screen-film System[J];Chinese Medical Equipment Journal;2008-09
6 ZHANG Xiao-jing,QIN Wei-chang,HUANG Ling,ZHAO Yong-xia(General Hospital of PLA,Beijing 100853,China);Test and Study of Performance Parameter of Digital Radiography System[J];Chinese Medical Equipment Journal;2010-04
7 TANG Feng1, ZHAO Lei2, WANG Peng-cheng1, XU Long-chun2 (1.Department of Radiology, Taishan Medical University, Taian Shandong 271016, China; 2.Imaging Center of Affiliated Hospital, Taishan Medical University, Taian Shandong 271016, China);The Effect of Changing Digital Angiography System Settings on Radiation Dosage and Image Quality[J];Chinese Journal of Medical Physics;2009-03
8 NIU Yan-tao, QI Wei-wei, ZHANG Fu-li, et al.Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730,China;;A comparative study of energy responses of computed radiography system and screen-film system[J];Chinese Journal of Radiation Mediation and Protection;2005-01
9 ZHAO Song, ZHANG Jun-yi, ZHANG Wei-min,et al. ~Department of Radiology,Seconded Affiliated Hospital,Zhejiang University, Hangzhou 310009,China;Low-dose investigation of digital radiography system image[J];Chinese Journal of Radiological Medicine and Protection;2006-04
10 JIANG Rong,HE Wen,WANG Ke-yang(Department of Radiology,Beijing Friendship Hospital Affiliated to Capital Medical University,Beijing 100050,China);Effect of contrast changing in digital chest radiograph in the diagnosis of costal fracture[J];Chinese Journal of Medical Imaging Technology;2007-11
【Secondary References】
Chinese Journal Full-text Database 2 Hits
1 TIAN Lan, GONG Jianping, JIANG Zhen,et al.Department of Radiology,The Second Affiliated Hospital of Soochow University,Suzhou,Jiangsu Province 215004,P.R.China;Application of Low Dose Photography in Pelvic Digital Radiography[J];Journal of Clinical Radiology;2011-02
2 ZHONG Zhao-hui1,LU Dong-sheng2(1.Department of Radiology,2.Department of Medical Engineering,Beijing Friendship Hospital,Capital Medical University,Beijing 100050,China);Impact of exposure parameters on chest image quality of mobile digital radiography[J];Chinese Journal of Medical Imaging Technology;2011-03
©2006 Tsinghua Tongfang Knowledge Network Technology Co., Ltd.(Beijing)(TTKN) All rights reserved