Full-Text Search:
Home|Journal Papers|About CNKI|User Service|FAQ|Contact Us|中文
《Journal of Instrumental Analysis》 2010-11
Add to Favorite Get Latest Update

Electrochemical Sensor for Diazepam Based on Molecularly Imprinted Film Modified Screen Printed Electrode

LIU Xiao-fang,YAO Bing,LIU Guo-yan,CHAI Chun-yan(School of Agriculture and Biology,Shanghai Jiao Tong University,Shanghai 200240,China)  
A molecularly imprinted film modified screen printed electrode(SPE) was prepared by cyclic voltammetry with electro-polymerization of o-phenylenediamine in the presence of diazepam which was removed later.The electrochemical characteristics of the imprinted and non-imprinted films were evaluated by differential pulse voltammogram.Potassium iodide was used as the probe between the modified electrode and the base solution,and a method for determination of diazepam was established.The molecularly imprinted film modified SPE showed the advantages of easy fabrication and replacement.The experimental results showed that the optimum incubation was 3 min,and a good linear relationship was obtained between peak current and concentration of diazepam in the range of 2.0×10-7-1.0×10-5 mol/L with a detection limit of 2.5×10-8 mol/L.The spiked recoveries of diazepam from meat sample ranged from 92% to 95%.The electrochemical sensor was successfully applied in the detection of diazepam in several samples.
【Fund】: 国家高技术研发发展计划(863计划 2006AA10Z439)
【CateGory Index】: R914
Download(CAJ format) Download(PDF format)
CAJViewer7.0 supports all the CNKI file formats; AdobeReader only supports the PDF format.