Determination of Aluminum, Titanium in Alloy Steel by Inductively Coupled Plasma-Optical Emission Spectrometry

Mi Hanbing (Physics & Chemistry Testing Laboratory of NO. 1, Measuring & Testing Center,Sichuan Changcheng Special Steel Co. Ltd., Pangang Group ,Jiangyou, Sichuan 621701)  
A method for determination of aluminum, titanium in alloy steel by Inductively Coupled Plasma-Optical Emission Spectrometry was presented. By ameliorating various aspects such as analysis spectral line, matrix effect, acidity and the measure conditions of instruments, the optimal conditions of determining aluminum, titanium with ICP-OES have been obtained in alloy steel: power of 1250W, nebulizer flow at 0.70L/min, auxiliary flow at 0.2 L/min, sample flow rate at 1.7mL/min. Under the best determination condition, the precision and accuracy and comparable tests for the samples were carried out. The proposed method has been used for conventional analysis with satisfactory results.
Download(CAJ format) Download(PDF format)
CAJViewer7.0 supports all the CNKI file formats; AdobeReader only supports the PDF format.
©CNKI All Rights Reserved