Design of the cold light source system for portable endoscope

ZHOU Wen-guang;KONG Yue;WANG Chun-fei;Chenggong Hospital,Xiamen University and The 174 Hospital of PLA;  
Objective: To design a portable endoscope cold light source system. Methods: Using special LED as a cold light source, optical fiber astransmission medium, optical system design LED to the front end of the endoscope; with TPS61165 high currentdrive chip, STM32 generates a PWM signal, the control circuit design of LED; cooling system through the external embedding mode design of LED lighting. Results: After clinical testing, cold light source illumination of the design of the system is stable, high color rendering index, color temperature low, good heat dissipation effect. Conclusion: The design of the lighting system is very suitable for theuse of portable endoscope.
【Fund】: 南京军区2013年度科技创新项目资助课题(MS087)“野战便携式内窥镜视频系统的研制”
【CateGory Index】: R197.39
Download(CAJ format) Download(PDF format)
CAJViewer7.0 supports all the CNKI file formats; AdobeReader only supports the PDF format.
【Citations】
Chinese Journal Full-text Database 5 Hits
1 WANG Jing,WU Fu-gen(College of Physics and Photo-electricity Engineering,Guangdong University of Technology,Guangzhou 510006,China);Key issues of improving cooling of high-bightness LED[J];Electronic Design Engineering;2009-04
2 ;The Discussion on Calibration Method of Medical Endoscope[J];Metrology & Measurement Technique;2012-01
3 LI Xiang-hong,TANG Wei,ZHANG De-chang(Digest Department of Southwest under Chongqing 3rd Military Medical University, Chongqing 400038, China);Working Principle and Maintenance of Electronic Endoscopy[J];Information of Medical Equipment;2007-07
4 CHEN Ji-ming,ZHANG Lian-qiang,TIAN Xiao-dong Medical Engineering Department,General Hospital of Airforce,Beijing 100142,China;Use and Maintenance of Electronic Endoscope[J];China Medical Devices;2010-05
5 CHEN Shan-qiong The Affiliated Hospital of Youjiang Medical Univerity for Nationality, Baise 533000, China.;Use, maintenance and common troubleshooting for FUJINON EPX-4400 endoscopic system[J];China Medical Equipment;2013-02
【Co-citations】
Chinese Journal Full-text Database 10 Hits
1 ZHANG Xiao-ling,WEI Xiu-ming(WuHu Ruby Automotive Lighting Systems Co.,Ltd,Wuhu 241000,China);The cooling and optical failure Of LED automobile headlamps[J];Journal of Anhui University of Technology and Science(Natural Science);2010-01
2 Wu Jun1,Li Shuzhi2,Yang Weiqiao2,Zhang Jianhua1(1.Key Lab.of Advanced Display and System Applications of Ministry of Education,School of Mechatronics Engineering and Automation,Shanghai University,Shanghai 200072,China;2.Shanghai ResearchCenter of Solid-State Lighting Engineering and Technology,Shanghai 201203,China);Study on Heat Sinks for High Power LEDs[J];Semiconductor Technology;2010-10
3 Wang Shuang1,Xiong Feng1,Yang Jiexiang2,Yu Tao1(1.School of Mechanical & Electrical Engineering and Automation,Shanghai University,Shanghai 200072,China;2.Semiconductor Lighting Engineering Research Center for Engineering ofShanghai,Shanghai 201203,China);Research on Optical Performance of Blue LED Floodlights Under High-Temperature Aging[J];Semiconductor Technology;2011-01
4 HAN Yuanyuan,GUO Hong(General Research Institute for Nonferrous Metals,Beijing 100088);Application of Numerical Simulation in Thermal Analysis of High Power LEDs Package[J];Materials Review;2010-21
5 LIU Hong1,ZHAO Qin1,JIANG Lanfang2,RUAN Lingwei 11.College of Mechanical Engineering,Zhejiang University of Technology,Hangzhou 310014,China;2.College of Zhijiang,Zhejiang University of Technologn,Hangzhou 310024,China;The Structure Design of Radiators for Integrated High-Power LED Street Light[J];Chinese Journal of Electron Devices;2010-04
6 HUA Yun-feng LI Zheng-xian DU Ming-huan DU Ji-hong WANG Yan-feng LI Hong-zhan (Northwest Institute for Nonferrous Metal Research.Xi'an 710016.China);Study on the Thermal Design of Tube Heat Exchanger for High Power LED[J];Optics & Optoelectronic Technology;2011-02
7 BAI Kun1,WU Ligang1,NIE Qiuhua1,DAI Shixun1,ZHOU Boyou2,MA Xiangjun1(1.College of Information Science and Engineering,Ningbo University,Ningbo Zhejiang 315211,China)(2.Ningbo Self Electronics Co.Ltd.,Ningbo Zhejiang 315103,China);Heat-release package design and analysis of high-power white LED down light[J];Optical Technique;2012-03
8 Zeng Su-qiong,Zhang Xue-cheng(School of Electronics and Information Technology,Jiaying University,Guangdong Meizhou 514015);High Efficiency and High Precision Control and Drive Circuit Design for LED[J];Electronics Quality;2013-08
9 Chen Yingcong;Wen Shangsheng;Wu Yuxiang;School of Materials Science and Engineering,South China University of Technology;State Key Laboratory of Luminescent Materials and Devices,South China University of Technology;College of Automation Science and Engineering,South China University of Technology;;Thermal Analysis for LED Chipon Board Package Based on Plastic Radiator without Substrate[J];Acta Optica Sinica;2013-08
10 ZHANG Miao;HUA Chuxia;LIN Zekun;Department of Automation ,Guangdong University of Technology;;An Optimization Design for High-power LED Lamp Radiator[J];Optoelectronic Technology;2014-01
China Proceedings of conference Full-text Database 1 Hits
1 ZHAO Min,CHEN Zhi-ping,ZHANG Ju-yong,LI Rong (College of Mechanical Engineering,Hangzhou Dianzi University,Hangzhou 310018,China);Thermal Analysis and Thermal Design Based on a High Power LED Lamp[A];[C];2011
【Secondary Citations】
Chinese Journal Full-text Database 10 Hits
1 QIAN Ke-yuan~1,ZHENG Dai-shun~1,LUO Yi~(1, 2)(1.Graduate School at Shenzhen,Tsinghua University,Shenzhen 518055,CHN;2.State Key Laboratory of Integrated Optoelectronics,Tsinghua University,Beijing 100084,CHN);Thermal Dispersion of GaN-based Power LEDs[J];Semiconductor Optoelectronics;2006-03
2 DAI Wei-feng,WANG Jun,LI Yue-sheng(Department of Material Science,Fudan University,Shanghai 200433,CHN);Transient Thermal Analysis of High-power LED Package[J];Semiconductor Optoelectronics;2008-03
3 Jiang Jie Yu Dao_yin Sun Zheng (School of Precision Instrument and Opto_electronic Engineering,Tianjin University,Tianjin 300072) (Opto_electronic Information Science and Technology Lab.,MOE);Real-time Image Processing System for Electronic Endoscope Based on FPGA[J];Journal of Electronic Measurement and Instroment;2001-04
4 Juanjuan Huang,Junhai Wen(Biomedical engineering department,Beijing institute of technology,Beijing100081,China);Working Principle and Status of Stereo Electronic Endoscope[J];Life Science Instruments;2009-03
5 MENG Jiang-Yun1,WEI Zhen-jun2,HUANG Qi-yang1,DU Hong1,WANG Xiang-dong1,LIANG Hao1 (1.General Hospital of PLA,Beijing 100853,China;2.General Hospital of Beijing Command of PLA,Beijing 100700,China);Retrospective analysis of the breakdowns and maintaining of Fujinon electronic endoscope[J];Biomedical Engineering and Clinical Medicine;2008-03
6 LI Weiguang,GE Zhizheng. Department of Gastroenterology,Renji Hospital,Shanghai Jiaotong University School of Medicine Shanghai Institute of Digestive Disease,Shanghai(200001);Autofluorescence Endoscopy[J];Chinese Journal of Gastroenterology;2007-07
7 HUANG Yaner,WU Jianyang(Zhongshan People's Hospital,Zhongshan 528403,China);The Principle and Prevention of Common Troubles for Electronic Endoscope[J];Clinical Medicine & Engineering;2012-02
8 WU Jing-xing,XIAO Bi-bo;Maintenance of the endoscope and its troubleshooting[J];Information of Medical Equipment;2005-03
9 WANG Jian-feng,DING Jiang-tao,YANG Yong-peng (Yantai Yuhuangding Hospital,Yantai Shandong264000,China);The maintainence and settlement of common troubles for OLYMPUS electronic endoscope system[J];Information of Medical Equipment;2005-06
10 GUO Yan;Routine Maintenance and Troubleshooting of PENTAX Electronic Endoscope[J];Information of Medical Equipment;2007-04