The Factors Influencing Peeling Strength of Adhesive Samples for Medical Devices

YANG Han;HAN Fei;LIU Hong-wei;WANG Xiao-wei;Shenzhen Key Laboratory of Drug Quality Standard Research;Shenzhen Institute for Drug Control(Shenzhen Testing Center of Medical Devical Devices);Jiangsu Province for Medical Equipment Testing;  
This paper analyzes the factors that influence peeling strength of adhesive materials for medical devices by changing different test conditions in accordance with the practical situation. It is concluded that the peel strength would be increased along with rising of the testing temperature, and the surface cleanliness of steel plate and the hold time after preparation of specimen are not brought too much influence on the peel strength.